FMB Oxford provides a full range of Cyberstar scintillation detector heads with various dimensions, crystal types, and thicknesses; allowing you to match a product to your efficiency, resolution, speed, and integration requirements.

Three types of scintillation crystal are available; NaI(TI), YAP and LaCl3.

The large area scintillation detector is the latest addition to the scintillation detector family.  It is primarily designed for use with a Bent Crystal Laue Analyser (BCLA), where the BLCA is positioned infront of the detector to discriminate the energy range before being counted by the detector.  The Detector consists of two photon multiplier tubes housed in an aluminium body with a Beryllium window.  The signals from the two internal detectors are counted seperately, this allows the position of the Detector head to be optimised in relation to the BLCA; when both signals are the same value the detector is in the optium position.

[/vc_column_text]

[vc_column_text animation_loading="no" animation_loading_effects="fade_in"]
Part number C48NA11B C48NA12B C48NA15B  C12NA12B
Body length 138mm 138mm 138mm 144mm
Body diameter  48mm 48mm 48mm 12.5mm
Detector aperture 30mm 30mm 30mm 6mm
Scintillator NaI(Tl) NaI(Tl) NaI(Tl) NaI(Tl)
Crystal thickness 1mm 2mm 5mm 2mm
Photomultiplier Ø 38mm 38mm 38mm 10mm
Photomultiplier gain 106 106 106 106
Window 0.2mm Be 0.2mm Be 0.2mm Be 0.2mm Be
Background noise <0.2cps @ 5keV <0.2cps @ 5keV <0.2cps @ 5keV <0.2cps @ 5keV
Minimum energy 1.5keV 1.5keV 1.5keV 1.5keV